GE Analytical Instruments, Inc. ©2005H Page 15 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
ENABLE/DISABLE
SET POLARITY
ENTER
CLEA
R
FIGURE THIRTEEN: Analyzer’s INPUT Menu
6.2.4. Select ENABLE/DISABLE. Press Enter.
DIGITAL INPUT CONTROL
CONTROL IS:
OFF
ENTER
CLEAR
A
RROWS TO CHANGE
ON
CLR = EXIT
CHANGE TO:
ENT = SET
FIGURE FOURTEEN: DIGITAL INPUT CONTROL screen
6.2.5. The digital input is either in a state of ON or OFF. To change this state, press
the up/down arrows to toggle between the two choices. When the state in the
change is correct, press Enter to set. To allow input, the state must be ON.
Press Clear to return to the INPUT Menu.
ENABLE/DISABLE
SET POLARITY
ENTER
CLEAR
FIGURE FIFTEEN: Analyzer’s INPUT Menu
6.2.6. Select SET POLARITY by using the up/down arrows, press Enter.
DIGITAL INP. POLARITY
RUN WHEN:
CLOSED
ENTER
CLEAR
A
RROWS TO CHANGE
OPEN
CLR = EXI
T
CHANGE TO:
ENT = SE
T
FIGURE SIXTEEN:
DIGITAL INP. POLARITY Screen
6.2.7. The DIGITAL INP. POLARITY is either in a OPEN or CLOSED condition. To change
this condition, press the up/down arrows to toggle between the two
conditions. When the condition is correct, press Enter to set. It is critical to
have the polarity set to the same condition as the switch, whether it is